SPSx-1µ 
Measurement solutions for moisture sorption and desorption analysis




Home
Products

Dynamic Vapor Sorption

SPSx-1µ_

Technical Data_


Contract Analyses
Publikations/Downloads
Vapor Sorption
News/Events
Contact
 
SPSx-1µ technical data / Version High Load (PDF 25 kB) / Version Advance (PDF 25 kB)


Version High Load Advance                                       
Laod Capacity:    23 Samples in sample pans Ø 33 mm
24 Samples in pans Ø 18 mm (not weighted)
or:    11 Samples in sample pans Ø 51 mm
24 Samples in pans Ø 35 mm (not weighted)
Max Load: 22g / 220g per sample (dual range) 22g per sample
Resolution: 1 µg / 10 µg (dual range) 1 µg
Repeatability: ± 5 µg /20 µg RMS (dual range) ± 2 µg***
Temperature Range: +10 °C. to +60 °C
Temperature Accuracy: over time ± 0.1 K
uniformity ± 0.5 K
Humidity range: 0 % r.H. to 98 % r.H.*
Dew Point: -70 °C. to +38 °C
Humidity Accuracy:
Long term stability
± 0,6 % r.F. (23°C ± 5°C)
better 1% per year
Dimensions Width: 488 mm
Depth: 630 mm
Height: 437 mm
(1024 mm with open cover)
Weight 68 kg **
Environmental Conditions:
Temperature: +15 °C. to +25 °C
Humidity: 75 % r.H. max.
Power Supply:
Voltage: 230 VAC ± 15 % 1⁄N 50 Hz
Power: 0,5 KW **
Gas Supply: Compressed air, Nitrogen
dried, oil-free
Operating Pressure: 1.5 bar to 10 bar
Dew Point: ≤ -70 °C
Gravimetric Validation DKD Calibration Certificate with “Min Weight” certification
Humidity Validation Salt validation test with humidity standards
Optional: Software validation package according to CFR21 part 11

*Note: The full humidity range is available at chamber temperatures slightly above room temperature.
**Note: Dimension, weight and power consumption do not include computer, keyboard and monitor.
***Note: An environment capable for the operation of an Ultra-Micro Balance is required.

continue>>>

  © PMT Analytical GmbH & Co. KG 2012