SPS23-100n 
Measurement solutions for moisture sorption and desorption analysis




Home
Products

Dynamic Vapor Sorption

SPS23-100n_

Technical Data_


Contract Analyses
Publikations/Downloads
Vapor Sorption
News/Events
Contact
 
SPS23-100n technical data (PDF 23 kB)


Laod Capacity:    23 Samples in sample pans Ø 18 mm
24 Samples in pans Ø 18 mm (not weighted)
Max load: 2000 mg per sample
Resolution: 0.1 µg
Repeatability: ± 1.5 µg RMS***
Temperature range: +10 °C. to +40 °C
Temperature accuracy: over time ± 0.1 K
uniformity ± 0.5 K
Humidity range: 0 % r.H. to 98 % r.H.*
Dew Point: -70 °C. to +38 °C
Humidity Accuracy:
Long term stability
± 0,6 % r.F. (23°C ± 5°C)
better 1% per year
Dimensions Width: 488 mm
Depth: 630 mm
Height: 437 mm
(1024 mm with open cover)
Weight 68 kg **
Environmental conditions:
Temperature: +15 °C. to +25 °C
Humidity: 75 % r.H. max.
Power Supply:
Voltage: 230 VAC ± 15 % 1⁄N 50 Hz
Optional: 110 VAC ± 15 % 1⁄N 60 Hz
Power: 0,5 KW **
Gas Supply: Compressed air, Nitrogen
dried, oil-free
operating pressure: 1.5 bar to 10 bar
dew point: ≤ -70 °C
Gravimetric validation DKD Calibration Certificate with “Min Weight” certification
Humidity validation Salt validation test with humidity standards
In Preparation: Software validation package according to CFR21 part 11

*Note: The full humidity range is available at chamber temperatures slightly above room temperature.
**Note: Dimension, weight and power consumption do not include computer, keyboard and monitor.
***Note: An environment capable for the operation of an Ultra-Micro Balance is required.